This cement hydration heat tester uses the dissolution heat method (based on Hess's Law: the heat effect of a chemical reaction depends only on the initial and final states of the system, not the reaction path) to determine cement hydration heat. The heat difference between unhydrated cement and hydrated cement dissolved in standard acid represents the hydration heat released at the specified hydration age. Suitable for hydration heat testing of moderate-heat Portland, low-heat slag Portland, ordinary Portland, slag Portland, pozzolanic Portland, fly ash Portland and other cement types at any hydration age.
Test Principle: Unhydrated cement samples and cement samples hydrated to specified ages are each dissolved in standard acid within the calorimeter. The dissolution heat is measured. The difference between unhydrated and hydrated cement dissolution heats represents the hydration heat released at the corresponding age.
Structure:
Vacuum flask: 650ml volume, 75mm inner diameter, 160mm height, excellent thermal insulation
Beckmann thermometer: Range 5~6℃, division 0.01℃, high precision
Constant temperature water bath: 20℃±0.1℃ for stable test conditions
Stirring device: Ensures thorough reaction of acid and sample
Calorimeter system: Precisely measures dissolution heat
Technical Features:
Hess's Law dissolution heat method for reliable and accurate results
Excellent vacuum flask insulation for minimal heat loss
Beckmann thermometer 0.01℃ division for high measurement accuracy
Water bath temperature control ±0.1℃ for stable conditions
Suitable for various cement types and any hydration age
Technical Parameters:
Power: 2500W / AC220V / 50Hz
Net weight: 100kg
Applications:
Cement manufacturers: Hydration heat quality control
Building material testing institutions: Cement hydration heat testing
Hydraulic engineering: Dam cement hydration heat control
Research institutions: Cement hydration mechanism research
Available in different specifications: different vacuum flask volumes (650ml, 750ml, etc.) and different temperature control accuracies (±0.1℃, ±0.05℃, etc.) to suit various cement types and testing standards.
Mainly used in cement manufacturers, building material testing institutions, hydraulic engineering quality testing stations, research universities, and more for cement hydration heat determination.